Axis BTA 2100D Instrukcja Operatora Strona 192

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 597
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 191
190
GC peaks) the maximum ionization time of 65 milliseconds is selected, which
offers maximum sensitivity. As the amount of analyte increases, the ionization
time is decreased automatically in order to prevent saturating the ion trap with
ions.
The ion signals in each microscan are scaled automatically to correct for the
variation in ionization time. Microscans are averaged to improve the spectral
quality.
Ion Formation
In the EI mode electrons are passed through the ion trap cavity during the
ionization period. These electrons collide with the neutral sample molecules
present and transfer some of their energy to the molecules. These energetic
molecules decompose through a series of reactions producing ions and neutral
fragments. This set of fragment ions, characteristic of the sample molecule,
makes up the mass spectrum. The degree of fragmentation is a function of the
internal energy of the molecules, which is a function of the energy imparted by
the electrons, as well as the ion trap temperature and the axial modulation
voltage amplitude.
The ionization time is the time (between 10 µs and 65 milliseconds) during which
the electron gate voltage is positive, and the electron beam from the filament is
permitted to enter the ion trap. The electron beam generates ions by colliding
with sample and/or background molecules which are present within the ion trap.
The AGC software provides control of the ionization time so that the quantity of
ions can be maintained at the optimum level.
The number of ions formed in the trap is also a function of the intensity of the
electron beam. In general, increasing the filament emission current increases the
quantity of ions produced. The relationship between the filament emission
current and the measured TIC (i.e., the number of ions generated in the ion trap)
is not a strictly linear one. In particular, the emission current and the measured
TIC do not vary linearly for emission current values above 40 A and at high
sample concentration.
Keep in mind that both the ionization time and the intensity of the electron beam
directly control the number of ions formed, and that the ion trap performs best
(producing the highest quality spectra with good resolution) when an optimum
quantity of ions (or “charge”) exists within its chamber.
AGC Prescan
The 0.1 millisecond ionization is followed by a ramping of the rf voltage to eject
the permanent background from 35 m/z to an upper mass limit selected by the
user. This upper mass limit is defined as the background mass. The default value
for the upper mass limit is 45 m/z; as a result, all masses below the background
mass (i.e., those masses corresponding to the chemical background) are ejected
from the trap.
The rf voltage is then ramped to its maximum value, and all ions with masses
above the background mass (i.e., those masses corresponding to sample ions)
are ejected from the ion trap and are detected by the electron multiplier. The
electron multiplier produces a signal proportional to the number of ions detected.
Przeglądanie stron 191
1 2 ... 187 188 189 190 191 192 193 194 195 196 197 ... 596 597

Komentarze do niniejszej Instrukcji

Brak uwag