336
Internal Standard Section
Item Description
Use
Yes or No
If Yes, an internal standard addition will be used.
When internal standard addition is used, the internal
standard solution will be drawn up into the syringe
from the specified solvent vial before the sample is
drawn up.
Vial I, II, or III
Selects which solvent vial contains the internal
standard.
Volume (uL) 0 to 9.0 microliters with 10 uL syringe
0 to 4.9 microliters with 5 uL syringe
Specifies the amount of internal standard to be drawn
into the syringe before the sample.
Drawup Speed (uL/sec) 0.1 to 50.0 microliters per second with 10 uL syringe
0.1 to 25.0 microliters per second with 5 uL syringe
Specifies the speed at which the internal standard
will be drawn up into the syringe.
Pause Time (sec) 0 to 9.9 seconds
Specifies how long the syringe is to remain in the
internal standard vial after drawing up the internal
standard.
Air Gap Yes or No
If Yes, 1 microliter of room air will be drawn into the
syringe to create an air gap between the internal
standard and the sample.