User Defined Parameters
Item Description
Solvent Flush Sampling When Yes is selected, the solvent plug size needs
to be specified and whether upper and lower air
gaps are to be used must be selected. When No is
selected, the solvent plug size and upper and
lower air gaps are turned off.
Syringe Wash Time (sec) 6 to 180 seconds
Sets the amount of time that the wash solvent will
be flushed through the syringe.
Air Dry After Wash This parameter may only be selected when Solvent
Flush Sampling is set to No. When Yes is selected,
the compressed air supplied to the 8200
AutoSampler will be blown through the syringe
after the wash cycle to dry the syringe. When No is
selected, air drying is not done.
Solvent Plug Size (ul) 0.0 to 3.0 microliters (10 microliter syringe)
0 to 30 microliters (100 microliter syringe
The volume of solvent loaded into the syringe used
to flush the sample from the syringe needle during
injection.
Vial Needle Depth (%) 0 to 100 percent
Sets the syringe needle depth in the vial. May be
used for doing ambient headspace or multiple
liquid phase sampling.
Uptake Speed (ul/sec) 1.0 to 5.0 microliters (10 microliter syringe)
10 to 50 microliters (100 microliter syringe
Sets the plunger speed to provide the specified
volumetric sample flow into the syringe.
Upper Air Gap
Lower Air Gab
When Yes is selected, room air is drawn into the
syringe to create the corresponding air gap. When
No is selected, the sampling step to create the
corresponding air gap is skipped.