Item Description
The noise used by Data Handling depends upon the noise
options set in the Integration Parameters Window in the Data
Handling section of the Method. The report includes the noise
value for this run, and the noise value actually used, if it is
different.
Manual
injection
(injection
information)
Indicates the type of injection device used (for example,
manual injection, 8200 AutoSampler, or AI-200 AutoSampler),
and the rack, vial and injection volume, if appropriate.
Run Log The Run Log is included in the report if the Run Log checkbox
is checked in the Results Format portion of the Report section
of the Method. The Run Log is the instrument control Method
that was used in the run that generated the Data File. For
3400 and 3600 GCs, it is the Method as it was downloaded
before the run began. The 3400 or 3600 Event Log (included
in the Run Log) will describe any changes that occurred after
run start. For all LC modules, the Run Log is the Method that
was actually executed on the LC module itself. Any changes
that occurred during the execution of the Method will be
incorporated into the Run Log. Also included with the Run Log
is information from the Module Information Log, which can be
edited in System Control on a per-module basis.
Time Events Data Handling timed events (programmed in the Time Events
Window in the Data Handling Section of the Method) are listed
with the times that they actually occurred.
Error Log The Error Log is included in the report if the Error Log
checkbox is checked in the Results Format portion of the
Report section of the Method. The Error Log is a list of errors
that occurred during the run that generated the Data File.
Calibration
Report
The Calibration Report is included in the report if the
Calibration Report checkbox is checked in the Results Format
portion of the Report section of the Method. The Calibration
Report is the text portion of the Calibration Block Report for
the Method used to calculate results. It includes the calibration
responses and curve coefficients for each peak.
Revision Log The Revision Log is included in the report if the Revision Log
checkbox is checked in the Results Format portion of the
Report section in the Method. The Revision Log contains the
date/time and Method name used every time a recalculation is
performed on the Data File. Results deletion is also
documented in the Revision Log. Documentation of
recalculations can be disabled from the Security
Administration application.
Original Notes The Original Notes are included in the report if the Notes
checkbox is checked in the Results Format portion of the
Report section of the Method. The Original Notes are the
notes set in the SampleList in System Control when the
injection was performed that generated the Data File.
Appended
Notes
The Appended Notes are included in the report if the Notes
checkbox is checked in the Results Format portion of the
Report section of the Method. The Appended Notes are the
notes set in the SampleLog (in System Control) or in the
Reintegration List (in Interactive Graphics) during the most
recent recalculation performed on the Data File.