Item Description
Deviation Tolerance 0.0 to 500
Sets the percentage of deviation from the calibration
curve beyond which a result is considered out of
tolerance.
Out-of-Tolerance Action No Action, Increment Error Count, Terminate Sample
List, Halt Automation
Sets which action is taken when results for a verification
run are out of tolerance. Either does nothing (No
Action), adds 1 to the error count (Increment Error
Count), stops the run and goes to the next SampleList
(Terminate Sample List), or halts automation (Halt
Automation) each time a verification fails.
Time Events Table
Item Description
Start 0 to 1440.0000
Sets the start time for the selected event.
Event WI, II, GR, VB, SR, FP, SP, HF, HB, HM, SN, TP, and PR.
Sets the event type to be added to the table. Choose to add
a width (WI), inhibit integrate (II), group (GR), solvent reject
(SR), valley baseline (VB) event, forced peak (FP), split
peak (SP), horizontal forward (HF), horizontal backward
(HB), horizontal minimum (HM), signal to noise ratio (SN),
tangent percentage (TP), or peak reject (PR) event.
Value/End Time Depends on event type
For II, GR, SR, VB, FP, HF, HB, and HM events, sets the
end time for the selected event. For WI, there are a discrete
set of values between .5 and 256 seconds. For SP, the
value is not applicable. For SN, the range is 1-256. For TP,
the range is 0-100%. For PR, the range is 0-1000000
counts.
Description Describes the legal values.
Add Adds the active parameters to the Time Events Table as a
new time event.
Insert Inserts a new row above the highlighted row.