If you indicate that Coolant is turned on, specify the desired “Enable Coolant at”
temperature and “Coolant Timeout”.
Use the spreadsheet to build a temperature ramp program to heat and/or cool the
1079 injector.
The first row of the spreadsheet contains the Initial temperature and hold time for
the 1079 injector. The 1079 will equilibrate to this setting when the method is
activated, and will be restored to this setting when each chromatographic run is
completed.
Rows after the first row of the spreadsheet contain the programmed settings for
each of the ramp segments. Each segment will ramp to the specified
temperature at the specified rate (assuming the rate is achievable), and then
hold the temperature for the specified time. (Note that the Rate in the first row is
always blank and cannot be edited. Also note that the entire Total column cannot
be edited.)
1079 Split Ratio
If the 1079 Injector is connected to a Type 1 EFC, press the “Split Ratio” button to
build the Split Ratio Time Program.
NOTE: You must have already configured the Type 1 EFC in the corresponding
position in the Flow/Pressure Section before you can program the split ratio.
Specifically, to build a 1079 Split Ratio Program, a Front 1079 must have a
corresponding Front Type 1 EFC, a Middle 1079 must have a corresponding
Middle Type 1 EFC, and a Rear 1079 must have a corresponding Rear Type 1
EFC.
Use the Split Ratio spreadsheet to build a split ratio time program to control the
1079 injector / Type 1 EFC combination.