Axis BTA 2100D Instrukcja Operatora Strona 247

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 597
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 246
245
More User Defined Settings Dialog Box
This dialog box is accessed from “More User Defined…” button that is displayed
when the User Defined mode is selected.
Solvent Plug Settings
Item Description
Vial I, II, or III
Selects which solvent vial to use for the solvent plug.
Volume (uL) 0 to 10.0 µL with 10 µL syringe
0 to 5.0 µL with 5 µL syringe
Specifies the amount of solvent to be drawn into the
syringe before the sample or internal standard.
Drawup Speed
(uL/sec)
0.1 to 50.0 µL per second with 10 µL syringe
0.1 to 25.0 µL per second with 5 µL syringe
Specifies the speed at which the solvent will be drawn up
into the syringe.
Pause Time (sec) 0 to 9.9 seconds
Specifies how long the syringe is to remain in the solvent
vial after drawing up the solvent.
Air Gap Yes or No
If Yes, 1 µL of room air will be drawn into the syringe to
create an air gap before the solvent plug.
Przeglądanie stron 246
1 2 ... 242 243 244 245 246 247 248 249 250 251 252 ... 596 597

Komentarze do niniejszej Instrukcji

Brak uwag