Software Parameters
Segment Setpoints
Scan Time
The duration of each scan (in seconds). The scan time can be set to any value
from 0.100 to 5.000 seconds per scan. Below the scan time input box is
displayed the number of microscans per scan for the selected mass range and
scan time. The number of microscans per scan is computed automatically; it
cannot be set directly. It is displayed for information only. The minimum duration
for the scan time is determined by the selected mass range.
Multiplier Offset
To improve detection limits, the multiplier voltage may be increased or decreased
by 300 volts of the Electron Multiplier 10
5
gain. The same multiplier gain must be
used for determining the calibration curve and analyzing unknown samples.
Emission Current
The number of ions formed in the trap is a function of both the ionization time
and the intensity of the electron beam. In general, increasing the filament
emission current increases the quantity of ions produced. The relationship
between the filament emission current and the measured TIC (i.e., the number of
ions generated in the ion trap) is not a strictly linear one. In particular, the
emission current and the measured TIC do not vary linearly for emission current
values above 40 µA and at high sample concentration.
In certain cases, a high emission current can be used to increase the sensitivity.
One of these is EI/MS/MS where the MS/MS isolation step eliminates the higher
background that would normally arise from a high emission current. The
emission current may be set from 5 to 100 µA.
Count Threshold
The minimum number of intensity counts per sampling interval required before a
signal is recorded. (The maximum number of counts per sampling interval is
4095.) The peak threshold can be set to any value from 0 to 1000 counts; it is
typically set to 0 or 1 count.