Edit calibration coefficients. Locked coefficients are not updated by calibration runs.
Verification Setup
Verification Setup options affect the behavior of verification runs. Refer to the on-
line help in Interactive Graphics for details on all Verification Setup fields.
Filling the Peak Table
The fastest way to build a Peak Table is to create one from the detected peaks in
the active chromatogram. To do so, pull down the Edit Method menu.
Select the Fill Peak Table command.
The Fill Peak Table window is displayed at the top of the screen, above the
Chromatogram Display Window.