Item Description
Define Peak
Windows...
Opens the Define Peak Windows dialog box to set time
windows for reference peaks and other peaks or to set an
Unretained Peak Time.
Print Prints the contents of the current peak table.
Define Peak Windows Dialog Box
For both types of peaks, you set an absolute time window and a percentage time
window. The actual time window used will be equal to the absolute time plus the
percentage time.
Item Description
Define Reference
Peak Windows:
Width
0.00 to 200.0
Sets an absolute time that will be added to and subtracted
from the retention time of a reference peak to define a time
window.
Define Reference
Peak Windows:
Retention Time %
0 to 100
Sets a percentage of the retention time of a reference peak
that will be added to and subtracted from the retention time
to define a time window.
Define Other Peak
Windows: Width
0.00 to 200.0
Sets an absolute time that will be added to and subtracted
from the retention time of a non-reference peak to define a
time window.
Define Other Peak
Windows: Retention
Time %
0 to 100
Sets a percentage of the retention time of a non-reference
peak that will be added to and subtracted from the retention
time to define a time window.
Unretained Peak
Time (minutes)
0.00 to 1440.00
Sets the Unretained Peak Time (the amount of time taken
by an unretained compound to pass through the
chromatograph) for the Method. Used in RRT calculations.