Item Description
Displays the actual temperature of the 1079 injector oven at
the 3800 GC.
Split State On or Off.
Displays the current split state of the 1079 injector method.
Split Ratio 1 to 10000 or Off
Displays the current split ratio of the 1079 injector method.
The split ratio is defined as the Column Flow + the Split Flow
/ the Column Flow.
Enable 1079
Injector Coolant at
30 °C to 450 °C
Displays the programmed injector oven temperature at which
the 1079 injector oven coolant is enabled.
Timeout 0.01 min to 999.99 min
Displays the programmed coolant time-out value in the active
method for the 1079 injector oven.
Timed Out Yes or No.
Displays Yes if the coolant time-out period has elapsed for
the 1079 injector oven.
1041/1061 Injector Status:
Item Description
Ready Yes or No.
The 1041/1061 injector oven is Ready (Yes) if the setpoints have
been achieved and stabilized. The 1041/1061 injector oven is Not
Ready (No) if readiness to start a run has not been achieved or if,
during a run, the oven becomes not ready or goes “out of
tolerance”.
Fault Yes or No.
Displays if a fault in the 1041/1061 injector component has
occurred.
1041/1061
Oven
On or Off.
Displays if the 1041/1061 injector oven component is turned off or
on at the 3800 GC. The 1041 and 1061 injectors are isothermal
injectors.
Setpoint 50 to 450 °C
Displays the 1041/1061 injector oven programmed temperature
setting in the active method.
Actual 50 to 450 °C
Displays the actual temperature of the 1041/1061 injector oven at
the 3800 GC.