263
Item Description
Methanizer Oven On or Off.
Temperature (C) 450 C.
PFPD Detector
Item Description
Detector Oven On or off.
Turns the PFPD oven on or off.
Electronics On or Off.
Indicates whether the installed PFPD will be used or
not.
Temperature (C) 50- 450 C.
Specifies the PFPD Temperature.
Square Root Mode On or Off.
Time Initial - 999.99 min.
Range 8, 9, 10.
Autozero Yes or No.
Set to yes at initial time means that the PFPD Autozero
is on continuous before the run starts. Autozero is
automatically disabled once the run starts.
Add Adds a line to the spreadsheet.
Insert Inserts a line above the currently selected row in the
spreadsheet.
Delete Deletes the currently selected row(s) in the
spreadsheet.