311
SPT Injector Status:
Item Description
Ready
Yes or No.
The SPT injector oven is Ready (Yes) if the setpoints have
been achieved and stabilized. The SPT injector oven is Not
Ready (No) if readiness to start a run has not been achieved
or if, during a run, the oven becomes not ready or goes “out
of tolerance”. Some reasons the GC may go Not Ready
during a run are: the temperature program is too aggressive,
the oven door is opened, the coolant runs out, the power line
voltage drops, etc.
Fault
Yes or No.
Displays if a fault in the SPT injector component has
occurred.
SPT Oven
On or Off.
Displays if the SPT injector oven component is turned off or
on at the 3800 GC.
Coolant On or Off.
Displays if the SPT injector oven coolant is enabled in the
active method.
Setpoint
-99 to 450 °C
Displays the SPT injector oven programmed temperature
setting in the active method.
Actual
-99 to 450 °C
Displays the actual temperature of the SPT injector oven at
the 3800 GC.
Enable SPT Coolant
at
30 °C to 450 °C
Displays the programmed injector oven temperature at which
the SPT injector oven coolant is enabled.
Timeout
0.01 min to 999.99 min
Displays the programmed coolant time-out value in the active
method for the SPT injector oven.
Timed Out Yes or No.
Displays Yes if the coolant time-out period has elapsed for
the SPT injector oven.
Type 1 Injector EFC Flow/Pressure Status:
Item Description
Ready Yes or No.
The Type 1 Injector EFC component is Ready (Yes) if the
setpoints have been achieved and stabilized. The Type 1 Injector
EFC component is Not Ready (No) if readiness to start a run has
not been achieved or if, during a run, the EFC becomes not ready
or goes “out of tolerance”. Some reasons the GC may go Not
Ready during a run are: the temperature program is too
aggressive, the oven door is opened, the coolant runs out, the
power line voltage drops, etc.
Fault Yes or No.