Hardware Dialog Box
Item Description
Sampling Device Specifies the size and type of syringe installed in the
8200 AutoSampler, either a 10 microliter liquid
sampling syringe, a 100 microliter liquid sampling
syringe, or a Solid Phase MicroExtraction (SPME)
fiber and holder.
Wait for Read In When checked, the AutoSampler waits for a
hardware Ready input signal before starting. When
unchecked, the hardware Ready input is ignored.
Invert Ready In When checked, the hardware Ready input logic is
inverted (a high signal is a true). When unchecked,
the default hardware Ready input logic is used (a low
signal is a true).
Configuration Shows the injector that the current configuration file is
setup for.
OK button Saves the entries and closes the dialog box.
Set Config… button Selects the configuration file to use. The
configuration file specifies AutoSampler parameters
to use when injecting into different injector
configurations.
Cancel Discards edits and closes the dialog box.