8200 Method Command Reference
8200 Configuration Window
Item Description
Carrousel Type 12 Vials or 48 Vials
Select the tray that is to be used on the 8200
AutoSampler when this Method is run.
End Time
GC Cycle Time (SPME mode)
0 to 640 minutes
Specifies the number of minutes that the 8200
AutoSampler will wait before starting the
sample preparation for the next injection when
Prep Ahead is checked.
Prep Ahead When checked, the 8200 AutoSampler will
begin preparing the sample for the next
injection before the instrument goes to ready.
When unchecked, the 8200 will wait for an
Instrument Ready before starting the sample
preparation for the next injection.
Liquid Sampling Modes Select from the various liquid sampling modes
described in the following table. When a liquid
sampling mode is chosen, the Liquid
Sampling Options subwindow is shown. For a
detailed description of the liquid sampling
modes and their corresponding 8200
AutoSampler parameters, please refer to the
8200 AutoSampler Operator’s Manual.
Mode Description
Standard Mode Default mode of operation. A solvent flush
sampling mode used for most samples.