Dissociation
Excitation Time: The excitation time is the time required for collision-induced
dissociation (CID) by ion excitation. The excitation time range is 0 to 1000 msec.
The default excitation time is 20 msec for both resonant and nonresonant
excitation.
Customize Resonant Method
NOTE: Be sure that you have specified the parent ion mass before opening the
Customize Parameters window.
Customize Parameters window is separated into three sections. Ionization,
Isolation and Dissociation.
Ionization
Ionization Storage Level: The m/z value at which the rf is maintained during
ionization and the coarse isolation step. The range is 35 to 160 m/z, with a
default value of 48 m/z.
Ejection Amplitude: the amplitude (in volts) of the ejection waveform during the
coarse isolation step. The default value will work for most ions. For certain ions
that are less stable, this value may need to be reduced in order to minimize the
amount of energy imparted to the parent ion during isolation. The range of the
ejection amplitude is from 0 to 60 volts, with a default value of 20 volts.
Isolation
Low Edge Offset: This parameter provides a fine adjustment for the isolation
window to optimize the ejection of the mass just below the parent ion mass. The
low edge offset range is 0 to 20 steps. The default value is 6.
Large positive values widen the isolation window towards low masses.