Item Description
Group This column is shown when the Show Peak Group Totals
checkbox is checked in the Results Format portion of the
Report Method section. The group number for each peak is
set in the peak table in the Data Handling Method section.
Status Codes A code or codes specific to a peak in the results table.
Possible status codes are
R: Reference Peak (designated in the peak table)
*: No result can be calculated, check the calibration curve.
+: More than one result can be calculated, the first solution is
shown in the results field. Check your calibration curve or use
a lower order curve fit (specified in the peak table).
V: Peak fails verification (verification runs only)
M: Missing peak (if Report Missing Peaks is set in the
Integration Parameters Window in the Data Handling section
of the Method)
C: Result out of calibration range, check the calibration curve
and the range tolerance setting in the Method.
S: Internal Standard Peak (designated in the peak table for
internal standard and normalized percent calculations only)
U: User-defined peak endpoint (the peak size is affected by an
endpoint that was manually placed in Interactive Graphics).
Total
Unidentified
Counts
The total number of counts accounted for by peaks not
identified in the peak table.
Detected
Peaks
The total number of peaks detected during integration. Peak
detection is affected by parameters set in the Integration
Parameters Window in the Data Handling Section of the
Method.
Rejected
Peaks
The total number of detected peaks that were rejected based
on values set in the Integration Parameters Window in the
Data Handling Section of the Method.
Identified
Peaks
The total number of detected peaks that were identified in the
peak table.
Standard Peak
Amounts
Set in the SampleList or SampleLog (in System Control) or the
Reintegration List (in Interactive Graphics). This is used with
Normalized Percent and Internal Standard calculations and
identifies the amount of the Standard Peak(s) (designated in
the Peak Table Window in the Data Handling Section of the
Method).
Multiplier Set in the SampleList or SampleLog (in System Control) or the
Reintegration List (in Interactive Graphics). The result for each
peak is multiplied by this value.
Divisor Set in the SampleList or SampleLog (in System Control) or the
Reintegration List (in Interactive Graphics). The result for each
peak is divided by this value.
Baseline
Offset
The distance the chromatographic signal is above or below
zero at the start of the run. Baseline offset is measured in
microunits (units are volts or absorption units, depending upon
the detector). The baseline offset is equivalent to the
amplitude of the first data point in the Data File.
Noise Noise is the peak-to-peak noise measured in microunits (units
are volts or absorption units, depending upon the detector).