
370
Item Description
poor operation but will not harm either the GC
or the AutoSampler.
AutoSampler, Syringe Speed 1 to 30 (cm/sec), default=30
The Injector Injection Speed parameters
define the rate at which the needle will
penetrate the injector septum and descend to
the set Injector Needle Depth.
AutoSampler, Needle Depth 0 to 100, default=100
The Injector Needle Depth parameters define
how far into the designated injector the needle
will penetrate. 100% causes the needle to be
completely inserted to its 2-inch length.
AutoSampler, Plunger Count Limit (1-100,000) – This parameter defines the
expected number of strokes the plunger is
expected to survive without significant
degradation. Once the entered value is
exceeded, a message will be displayed on the
CP-8400/8410 Status page indicating the
plunger counter has exceeded the set value.
The default number of strokes is 10,000.
AutoSampler, Enable Buzzer Buzzer at Start of Cycle (On/Off) – A sound
will be generated to warn of impending tower
movements when this parameter is set to On.
The default value is On.
AutoSampler, Calibrate
AutoSampler
Displays the 8400/8410 AutoSampler
Calibration Dialog Box.
Calibrate Autosampler Dialog Box
This dialog box helps the user to calibrate the 8400/8410 AutoSampler. Items
that are not active are "grayed" in the actual screen.
Komentarze do niniejszej Instrukcji